MRI SAR Estimator

Specific Absorption Rate (SAR) measures power deposited per unit mass in tissue (W/kg). IEC 60601-2-33 limits whole-body SAR to 2 W/kg (normal mode) and local head SAR to 3.2 W/kg. SAR scales with B₁⁺ amplitude squared, frequency squared, and tissue conductivity.

Equations & Parameters ▸
\(\text{SAR}=\dfrac{\sigma|\mathbf{E}|^2}{2\rho}\approx\dfrac{\omega^2\mu_0^2 B_1^{+\,2}\sigma\,r_{eff}^2}{2\rho}\quad[\text{W/kg}]\quad\mu_0=4\pi\!\times\!10^{-7}\text{ H/m}\)
B₁⁺RF excitation field amplitude at the tissue (µT). Typical body coil: 10–20 µT for a 90° pulse.
fLarmor frequency. SAR scales as f² (or B₀²).
σTissue electrical conductivity (S/m). Muscle ≈ 0.7–0.9 S/m at 128 MHz.
ρTissue mass density (kg/m³). Muscle ≈ 1050 kg/m³.
Duty cycleFraction of time RF is on. SAR is time-averaged.
SAR formulaSAR = σ·|E|²/(2ρ) ≈ (ω·µ₀·B₁⁺)²·σ/(2ρ·µ₀²) = ω²·(B₁⁺)²·σ/(2ρ).
Physical constants used
µ₀4π×10⁻⁷ H/m
¹H γ/2π42.577 MHz/T (Larmor frequency per Tesla)
¹H at 1.5 T63.87 MHz
¹H at 3 T127.74 MHz
¹H at 7 T297.7 MHz
σ_muscle @ 128 MHz≈ 0.77 S/m
ε_muscle @ 128 MHz≈ 58
IEC SAR limit (WB normal)2 W/kg (over 6 min)
IEC SAR limit (head)3.2 W/kg (over 10 min)
Inputs
µT
peak RF field at tissue
MHz
¹H: 64/128/298 MHz at 1.5/3/7T
S/m
enabled for custom tissue
kg/m³
enabled for custom tissue
(0–1)
time-average fraction
Results

SAR

Peak local SAR
Time-averaged SAR
IEC whole-body limit (2 W/kg)
IEC head limit (3.2 W/kg)

Context

E-field estimate
ω·B₁⁺ product
B₁⁺ for 2 W/kg limit
Safety margin
⚠ Disclaimer: This is a simplified local SAR estimate for educational purposes only. Clinical SAR compliance requires full electromagnetic simulation with patient-specific models. Always consult IEC 60601-2-33 and vendor-provided SAR models for actual scanner operation.